background image

Model 3178

designated as 

slave 

instruments. The 

Oscillator In 

terminal of each 

slave 

unit (at the

instrument I/O connector) provides a connection point to the 

Oscillator Out 

termi-

nal of the 

master 

unit. The 

Oscillator Out 

and 

Oscillator Disable 

terminals of the

slave 

units are jumpered  to  disable  the  oscillator  internal  to  these  units.

Calibration Circuit. 

 A 

Cal Sense 

line is returned from the +

Signal

 connection

at  the  transducer  to  provide  for  shunting  the 

Cal Resistor 

across  a  bridge  leg  to

simulate an 

Equivalent Input.

 When the +CAL button is pressed, the 

Cal Resistor

is  shunted  across  the  +

Signal

  and  +

Sense 

connections.  This  action  occurs  by

grounding  the  negative  input  of  a 

Comparator 

through  the  +CAL  switch.  The

Comparator 

output  then  closes  an  analog  switch  to  which  the 

Cal Resistor 

is

connected.  The  connection  to  the  +

Sense 

line  is  then  made  through  normally-

closed contacts of the –CAL switch. Similarly, when the –CAL button is pressed,

the 

Cal Resistor 

is  shunted  across  the  +

Signal 

and  –

Sense 

connections.  The

Comparator

 is exercised as when the +CAL button is pressed, and the 

Cal Resistor

is switched to the –

Sense 

line  through  the  normally-open  contacts  of  the  –CAL

switch.

Calibration  in  the  positive  realm  can  also  be  remotely  checked.  When  the

Remote Cal 

input at the 3178 I/O connector is brought to a zero-volt (ground) level

through  the  action  of  an  external  switch,  transistor  driver,  etc.,  the  same  action

occurs  as  when  the  +CAL  button  is  pressed.

A line from the 

Cal Resistor 

is also provided for use with Lebow 

1600 Series

rotary transformer torque transducers. This type of transducer has an integral 

Cal

Resistor 

installed by the manufacturer. The line provided from the Daytronic 

Cal

Resistor 

enables the use of the Lebow 

Cal Resistor 

without  removing  the  resistor

integral to the 3178.

Signal Conditioner. 

  The +

Signal 

and –

Signal 

inputs from the transducer are

applied to a differential ac signal amplifier with excellent common-mode rejection.

Summed  into  the  amplifier  are  the  appropriate  resistive  and  capacitive  nulling

voltages from the R and C BALANCE controls. These enable the operator to set the

3178  output  to  zero  under  transducer 

no load 

conditions.

A high-pass filter/amplifier follows the input differential amplifier to provide

the  required  gain  and  eliminate  60-Hz  ripple  or  noise  related  to  the  rotational

frequency of the device when a rotating transformer torque transducer is used. The

Coarse 

SPAN  control  is  situated  at  the  output  of  the  amplifier  to  provide  a  gain

control.

16

Summary of Contents for 3000 series

Page 1: ...MODEL 3178 STRAIN GAGE CONDITIONER SB 5 1 INSTRUCTION MANUAL 3000 Instrument Series...

Page 2: ...reprinted reproduced or used in any form or by any electronic mechanical or other means including photocopying and recording or in any information storage and retrieval system without permission in wr...

Page 3: ...aired with CAL SENSE unconnected at Connector A SHIELD Daytronic 400 Series Transducer Connections 1 A 2 B 3 4 C 9 D A B C EXC EXC SENSE SENSE SIG SIG CAL SENSE B A Extra Wire paired with CAL SENSE un...

Page 4: ...ction 4 Wire Configuration for Cables Shorter Than 20 Feet SENSE EXC SENSE SIG SIG CAL SENSE 1 A 2 B 3 C 4 9 A SHIELD Extra Wire paired with CAL SENSE unconnected at Connector A EXC 6 Wire Configurati...

Page 5: ...DEL 3178 STRAIN GAGE CONDITIONER INSTRUCTION MANUAL Model 3178 Instruction Manual v SB 5 1 Pub No 3178M 5 1 Issued 03 01 Part No 91121 Daytronic Corporation Dayton OH Tel 800 668 4745 www daytronic co...

Page 6: ...This page intentionally blank...

Page 7: ...onditioner 1 2 Instrument Mounting Dimensions 4 3 Instrument Panel Mounting 5 4 I O Wiring 9 5 Front Panel Description 12 6 Block Diagram 17 7 Star Bridge Construction 20 LIST OF TABLES Table Page 1 S...

Page 8: ...ssary balancing and calibration controls and conditions amplifies the applied input to a standard Five Volt Data Signal Level which is the output analog signal level of 3000 Series Instruments Two ana...

Page 9: ...e Analog Outputs Two analog outputs available 0 to 5 volts with 50 overrange 5 milliamperes maximum Bandpass is dc to 2 Hz or dc to 400 Hz Active low pass filters provide for rolloff of 60 dB per deca...

Page 10: ...cteristics Positive realm calibra tion can be remotely checked by means of Remote Cal terminals on the instrument I O connector The 3178 Strain Gage Conditioner is also available in two additional for...

Page 11: ...Model 3178 A Bench Mounting B Rack Mounting C Panel Mounting Figure 2 Instrument Mounting Dimensions 4...

Page 12: ...Daytronic Corporation Figure 3 Instrument Panel Mounting 5...

Page 13: ...the front panel screwdriver adjustments and buttons are accessible These screws give approxi mately 1 8 inch of adjustment consequently this is the maximum panel thickness which should be used CAUTION...

Page 14: ...d even though the transducer calibration data mentions some other resistance value In Section 3 of this manual the tech niques described above are demonstrated If however the installed value of calibr...

Page 15: ...tor Figure 4 indicates three methods of remotely entering the calibration mode external switch transistor or TTL source The Remote Cal function provides a convenient method of periodically monitoring...

Page 16: ...I600 Fig 4 I O Wiring Data See the CORRECTION to Fig 4 in the front of this manual...

Page 17: ...I600 I600 I600 I600 I600 I600 I600 Fig 4 cont d See the CORRECTION to Fig 4 in the front of this manual...

Page 18: ...ALANCE control to obtain the minimum least positive or most nega tive output reading In some instances an integral digital indicator will be used to display the conditioner output Model 3278 or 3378 W...

Page 19: ...rols SPAN Controls The Coarse and Fine SPAN controls are used to set the output to the dead weight value when dead weight calibration is used They can also be used to set the output to the Equivalent...

Page 20: ...h the transducer loaded as in step f adjust the Symmetry control just to the right of the CAL button to obtain a dead weight reading equal to that obtained in step e If dead weight calibration is not...

Page 21: ...ide switch is used to turn ON primary power Overload protection is provided by a 0 25 ampere fuse mounted near the ac connection point When the slide switch is ON primary power is applied to the power...

Page 22: ...the output of the Integrat ing Amplifier changes the resistance of an Automatic Gain Control element FET to return the oscillator amplitude to its nominal value of 2 volts ac The Integrating Amplifier...

Page 23: ...ontacts of the CAL switch Calibration in the positive realm can also be remotely checked When the Remote Cal input at the 3178 I O connector is brought to a zero volt ground level through the action o...

Page 24: ...This page intentionally blank...

Page 25: ...Fig 6 Block Diagram...

Page 26: ...B within a decade of the cutoff frequency 5 VERIFICATION OF NORMAL OPERATION It is the purpose of this section to aid the user in determining in the event of a malfunction to which the Model 3178 is s...

Page 27: ...a transducer and cable known to be in good condition or by simulating a balanced transducer using either a commercially available transducer simulator or the simple star bridge arrangement shown in F...

Page 28: ...ason for this method of construction and it should be followed Connect the substitute or simu lated transducer to the instrument I O connector using a short 4 wire cable configuration as shown in Figu...

Page 29: ...This page intentionally blank...