DS-100C Operating Manual
page B41
ERROR
32
“INPUT LESS THAN OR EQUAL ZERO VOLTS”
The take I-V curve command cannot continue because the PV system
voltage is less than or equal to 0 V. Check PV system and DS-Tracer
disconnect switches. This error can also be caused by a reverse polarity
on the DS-Tracer to PV system connection.
ERROR
40
“DISCONNECT SWITCH IS OFF”
The DS-Tracer is trying to take an I-V curve, but the DS-Tracer
DISCONNECT switch is off.
ERROR
50
“INVALID NUMERIC PARAMETER”
Parameter must represent valid numeric value.
ERROR
60
“INVALID ANALOG I/O CHANNEL #”
The DS-Tracer accepts channel numbers 0 through 13.
ERROR
63
“DSP ERROR”
DS-Tracer DSP board has an error. Usually an A/D has failed or the
connect to the DSP board has failed.
ERROR
“UNKNOWN ERROR”
Some type of unknown error has occurred. Check parameters for proper
syntax.
Summary of Contents for DS-100c