DT1310 xx xx
20
20220128-V0
7.2. Connecting wire setting
Function
The Pt / resistance / potentiometer can be connected to the
input either with 4 / 3 / 2 wire.
The 4 wire connection gives the best result: the instrument
is able to compensate the connecting cable resistances.
The 3 wire mode is a compromise it may give a good result
when all the 3 wire have same resistance.
The two wire mode is the cheapest mode, use only when
there is no other possibility.
[Factory default: 4 wire.]
1. Select the appropriate connecting wire mode by clicking
the button.
The figure shows the factory default setting.
Sequence of operations
Warning!
Select always appropriate wiring mode.
Inappropriate selection may cause measuring error.
Summary of Contents for DT1310 Series