Colibrick A/D Converter
3 Installation
3.4.1 Measuring on multiple instruments
When measuring on multiple instruments using only a single
Colibrick
to
gather data, rather than dragging the entire
Colibrick
icon from the
Setup
Control Modules list
, drag the individual detector signals (
FID
,
UV-VIS
in
this case) to separate Instruments
. See
. The Instrument
to which the particular signal is connected is listed to the right
of the
detector name in the
Setup Control Modules
section.
Fig 8: System Configuration – multiple instruments
Note:
Each independent
Clarity Instrument
must have separate
Ext. Start
Dig. Input
and
Ready Dig. Output
pins configured according to the actual
wiring.
Note:
A larger number of Instruments can be set than the amount you have
purchased. You will not be able to measure on the surplus Instruments
(indicated by a blue symbol of the curve on the tab), but you may use them
e.g. for “offline” preparation of methods.
- 11 -