5 Troubleshooting
5.2.4 Data Acquisition - Simulated
▌ The title "Simulated" is displayed in the
window.
Fig 35: Data Acquisition Simulated
Description:
The corresponding Instrument only displays the simulated curve (from the file
CHANNX.DTA file) in the
window, or in other words has allocated
a
DEMO detector
on itself.
Solution:
Open the
dialog from the
window using the
System -
Configuration...
command and check the tab of the corresponding Instrument -
Instrument X
. If it only has the detector signals from the
DataApex DEMO
detector, it is necessary to reconfigure it. You can find more information how to add
a detector in the chapter
If you have a proper detector configured on your Instrument, the
Clarity
may be in
the
DEMO MODE
due to improperly entered user code. This is indicated by the
DEMO - TRIAL EXPIRED
red label in the main window. If so, check the user code
using the
Help
-
User Code
menu in the
Clarity
main window.