Chapter 3
42
Figure 13: Connecting Counter/Timer Signals for an Event Counting Operation
Without Using a Gate
Connecting Signals for Frequency Measurement
One way to measure frequency is to connect a pulse of a known duration to the Counter 0
Gate signal, as shown in
. In this case, the frequency of the Counter 0 In signal is the
number of counts divided by the period of the signal connected to the Counter 0 Gate input.
Figure 14: Connecting Counter/Timer Signals for a Frequency Measurement Operation
Using an External Pulse
Signal
Source
Counter 0 In
Ground
18
DT9816 Series
Module
19
Counter 0 In
(Number of pulses counted
during gate period)
Counter 0 Gate
(Determines period
for count)
Signal
Source
Ground
18
DT9816 Series
Module
19
16
Known Signal
Source