Operation ■ Changing Programmer Status
—68—
Data I/O • 096-0461-001G
back
Changing Programmer Status
PSV7000 may automatically disable a programmer for various rea-
sons, for example, if there are repeated continuity errors or vacuum
issues. The operator may also choose to disable a programmer, or
otherwise change the status of a programmer.
Disabling Programmers
If programmers are not performing and the error cannot be cleared,
disable those programmers and continue to use the other sites to pro-
cess devices.
To disable a programmer:
1.
At the
Run
window, click
Pause
.
2.
Right-click on the target programmer and select
All => DIS-
ABLED
. All sockets on this programmer change to black and are
effectively disabled.
Note:
To disable a single socket, right-click the socket and select
=> DISABLED.
The socket changes to black and is disabled.
To resume a job click
Run
. When the job finishes, click
Clear
or manu-
ally remove any devices from the disabled programmer sockets.
Clearing a Disabled Status
If a programmer is disabled (the PNP head skips it) it may be
re-enabled if there are no devices in it as follows:
1.
At the
Run
window, click
Pause
.
2.
Right-click on the target programmer and select
All => EMPTY
.
This DOES NOT REMOVE DEVICES FROM THE SOCKETS,
only changes the status.
Note:
To re-enable a single, empty socket, right-click the socket and
select
=> EMPTY.
This DOES NOT REMOVE DEVICES FROM THE SOCKET,
only changes the status.
All sockets on this programmer change to grey and are
re-enabled UNLESS THEY WERE NOT EMPTY.
3.
To resume a job click
Run
.
The programmer status
menu is shown in
Figure
49 on page 69.
An optional method of
changing programmer
and socket status is with
the
Programmer
Interface
window.
(Setup window > Sys-
tem > Programmers)
See the on-screen Help
for more information.