Piranha ES-Sx User Manual
03-032-20162-00
Teledyne DALSA
36
3.4 Saving and Restoring Settings
Saving and Restoring Factory and User Settings
Figure 7: Saving and Restoring Overview
Factory Settings
You can restore the original factory settings, including the factory calibrated pixel
coefficient set, at any time using the command
rfs
.
User Settings
There are two main sets of user settings: Area Mode user settings and TDI Mode user
settings. After issuing the user settings save command,
wus
, settings are saved depending
on which mode the camera is operating in when the command is issued. Also, when
operating in TDI Mode, digital gain and offset, and background subtract values are saved
as distinct values for Forward and Reverse directions. In other words, you can program
the camera to operate with a digital gain value of +5db in Forward direction and an
digital gain value of +3db in Reverse direction. Forward and Reverse direction settings
are saved simultaneously with the
wus
command. Note that when you switch directions,
the settings saved for that direction are automatically loaded.
Figure 8: How User Settings are Stored in the ES-Sx Cameras after issuing the wus Command
Area Mode
User Settings
All settings
saved for Area
Mode only
Camera is operating
in TDI Mode
Camera is operating
in Area Mode
User issues
command
wus
START
TDI Mode
User Settings
All settings, except
pixel coefficients,
saved for TDI
Mode only.