Piranha ES-Sx User Manual
Teledyne DALSA
03-032-20162-00
51
Error Messages
Camera Response
Comment
Error 01: Internal error xx>
Where xx is a code list below.
Only output during power up.
Customer should contact Teledyne DALSA customer
support.
Error 02: Unrecognized command> Command is not valid.
Error 03: Incorrect number of
parameters>
Too many or too few parameters.
Error 04: Incorrect parameter value> This response returned for
Alpha received for numeric or visa versa
Float where integer expected
Not an element of the set of possible values.
E.g., Baud Rate
Outside the range limit
Error 05: Command unavailable in
this mode>
E.g. SSF when in SEM 3
Error 06: Timeout>
Command not completed in time. E.g. CCF in SEM 3
when no external EXSYNC is present.
Error 07: Camera settings not saved> Indicates that user settings have been corrupted by
turning off the power while executing the WUS
command. Must build up new settings from factory
and re-save with WUS.
Error 08: Unable to calibrate - tap
outside ROI>
Cannot calibrate a tap that is not part of the end of line
statistics.
Error 09: The camera's temperature
exceeds the specified operating
range>
Indicates that the camera has shut itself down to
prevent damage from further overheating. (flashing
red)
Shuts down at internal temperature of 75˚C and will
not restart until below 65˚C (equivalent to 50˚C at front
plate).
Error 10: FPGA Flash Program Failed FCS failed either because of communication error or a
bad file was sent.