Dakota Ultrasonics
34
6) Press the
OK
key to set the
GATE1 WIDTH
and return to the menu screen,
or
ESC
to cancel entering the
GATE1 WIDTH
.
Note:
This sets the overall width of the gate in terms of the horizontal axis or
depth. Sets the right boundary of the gate, or where it turns off.
Set Threshold1 Level
1) Press the
UP
and
DOWN
arrow keys until
THRESHOLD1
is highlighted,
followed by pressing the
LEFT
and
RIGHT
arrow keys to scroll the value.
When the correct threshold level is displayed, proceed to the next setting.
2)
Alternatively
, press the
ENTER
key to display the
Digits Edit Box
.
3) Press the
UP
and
DOWN
arrow keys to scroll the highlighted value.
4) Press the
LEFT
and
RIGHT
arrow keys to scroll the digit locations.
5) Repeat steps
3 & 4
until the threshold level is correctly displayed.
6) Press the
OK
key to set the
THRESHOLD1
level and return to the menu
screen, or
ESC
to cancel entering the
THRESHOLD1
level.
Note:
This sets the overall sensitivity level of the gate which is the threshold.
Select Gate Polarity
1) Use the
UP
and
DOWN
arrow keys to scroll through the sub menu items
until
POLARITY1
is highlighted.
2) Press the
LEFT
and
RIGHT
arrow keys to toggle the
POLARITY
feature
positive or negative.
3) Press the
MEAS
key to return to the measurement screen.
Note:
The
GATE1
and
WIDTH1
values can now be adjusted or tuned directly
from the Hot Menus on the main measurement screen.