Dakota Ultrasonics
96
4) Press the
MEAS
key to return to the measurement screen in standard
portrait view.
5) Press the
OK
key to rotate the screen to
LANDSCAPE
view.
Note: Portrait and Landscape modes are quickly toggled by using the
OK
and
MEAS
keys. If a measurement mode with multiple values is enabled, like
PECT, more than one measurement value is typically displayed in portrait
mode. However in
LANDSCAPE
view, only one value will be displayed to
avoid too much interference with the A-Scan waveform. Pressing the
OK
key multiple times will toggle through all of the measurement available for
display.
11.11 DIM
The
DFX-7
has a built-in
DIM
feature to manage power more effectively. This feature
has adjustable time durations, until the power of the display is dimmed and current
draw reduced. The timer is constantly reset while the user is making measurements,
and is only activated when the gauge is idle for the duration of time the
DIM
value is
set for. This feature does not override the preset 5 min idle power off feature in any
way. However, if the gauge is turned on, set for a DIM of 120sec, and then left idle
for 5 minutes, the
DFX-7
will dim in 2 minutes, then power off in 5 minutes. The
procedures for adjusting the
DIM
time are outlined below:
Setting a DIM time