Cygnus 4+ Operating Manual
M5-CYG4P-M-01_Iss5.doc
30
The Resolution is set to 0.05mm, therefor the thickness
measurement must not change by more than +0.05mm or
-0.05mm for 2 seconds or more to be “stable”.
Measuring Small Diameter Pipe & Tubes
When measuring small diameter pipe and tubes with a twin
element probe, 75mm (3”) or under, you must ensure the face of
the probe is correctly aligned to the curvature of the pipe
otherwise measurements may be inaccurate.
The dividing line on the face of the probe should be at
right angles to the length of the pipe.
Display Hold Function
The thickness measurement can be ‘Held’ or frozen by simply
pressing the X/Cancel key while in any measurement screen.
1.
A thickness measurement is
displayed
Pipe
Probe Face
Probe Face Division is at
right angles to the pipe