Cygnus M5-C4P
Technical Specifications
Measurement Ranges
Single element 0° probes in Multiple Echo mode;
S2C/D probe
3 to 250 mm
[0.120 in. to 10.00 in.]
S3C probe
2 to 150 mm
[0.080 in. to 6.000 in.]
S5C/A probes
1 to 50 mm
[0.040 in. to 2.000 in.]
Twin element probes in Single Echo mode;
T2C probe
2.5 to 250 mm
[0.098 in. to 10.00 in.]
T5B probe
1.5 to 200 mm
[0.059 in. to 7.900 in.]
T7A probe
0.8 to 50 mm
[0.031 in. to 2.000 in.]
Twin element probes in Echo-Echo mode;
T2C probe
5 to 50 mm
[0.200 in. to 2.000 in.]
T5B probe
4 to 50 mm
[0.160 in. to 2.000 in.]
T7A probe
3 to 25 mm
[0.120 in. to 1.000 in.]
Probe Zero
Twin element probes are zeroed in the air without the need of
a zero block. The user is prompted to zero the probe at power
up and after a probe change.
Single element probes in multiple echo mode do not need to
be zeroed.
Measurement Modes
Single Echo mode with Twin element probe. Measures
thickness using the time from the delay line (zero point) to the
first back-wall echo. (UT Mode 2)
Echo-Echo mode with Twin element probe. Measures thickness
using the time between the first and second back-wall echoes.
(UT Mode 3)
Multiple Echo using Single element probe. Uses three matched
and verified back-wall echoes to determine the material
thickness and ignore surface coatings.
Resolution
Single Echo and Echo-Echo measurement modes;
Low
0.1 mm
[0.005”]
Medium
0.05 mm
[0.002”]
High
0.01 mm
[0.001”]
Multiple Echo measurement modes;
Low
(measurement > 120 mm)
0.1 mm
[0.005”]
Medium
(measurement < 120 mm)
0.05 mm
[0.002”]
Accuracy
±0.1 mm (±0.004”) or 0.1% of thickness measurement whichever is
the greatest.
Display
Type of Display
Colour TFT LCD with LED Backlight (front)
Display Size
240 x 128 Pixels QVGA. 2.4“, 47 mm (W) x 37 mm (H)