72
EXCEL V+ OPERATOR MANUAL
D2325, REV A, 2/19
Calibration Instructions
1.
Set up the laser energy meter in a convenient location so that the sensor head can be
easily reached with the handpiece. Set the meter display unit to the ENERGY mode.
2.
Ensure that all personnel in the room are wearing the appropriate eyewear.
3.
Connect the service computer to the serial port.
4.
Start the system in service mode and using the Cutera service software, place the system
in CALIBRATION mode.
5.
Follow the prompts on the calibration screen.
6.
Place the system in User mode and verify the calibration using an external meter.
End of Life Disposal - Environmental Information
The excel V+ system must be disposed of according to local laws and hospital practices. This product is
considered electronic equipment and must not be disposed of as unsorted municipal waste and must be
collected separately. Please contact the manufacturer or other authorized disposal company to
decommission your equipment.
Proper disposal of electronic equipment is required according to EU Directive 2002/96/EC Waste
Electrical and Electronic Equipment (WEEE).
Warranty Information
For specific and detailed warranty information for the excel V+
system, please refer to the first page of
your purchase “Agreement” and the last page of the “Terms and Conditions of Sale”.
CAUTION
Any work performed by unauthorized personnel will void all
warranties.
Summary of Contents for Excel V+
Page 1: ...I Operator Manual...
Page 8: ...viii EXCEL V OPERATOR MANUAL D2325 REV A 2 19 THIS PAGE IS LEFT BLANK...
Page 14: ...6 EXCEL V OPERATOR MANUAL D2325 REV A 2 19 Remote Interlock Plug Footswitch...
Page 104: ...A 2 EXCEL V OPERATOR MANUAL D2325 REV A 2 19 THIS PAGE IS LEFT BLANK...
Page 110: ...C 2 EXCEL V OPERATOR MANUAL D2325 REV A 2 19 THIS PAGE IS LEFT BLANK...
Page 112: ...D 2 EXCEL V OPERATOR MANUAL D2325 REV A 2 19 THIS PAGE IS LEFT BLANK...