45
EXCEL V+ OPERATOR MANUAL
D2325, REV A, 2/19
Adjusting the Temperature of the Sapphire Cooling Window
The CoolView handpiece features an integrated sapphire cooling window, which allows you to adjust
epidermal cooling to increase safety and efficacy. To adjust the temperature of the sapphire cooling
window:
1.
Press the
(temperature) button on the CoolView Treatment Screen to display the
temperature pop-up window.
2.
Select the desired temperature setting from the temperature pop-up window.
If you are in READY mode when you change the temperature setting, the system will revert to STANDBY
mode. When you press the
(laser status) button to toggle to READY mode, the window will cool
to the selected temperature.
NOTE
The selected temperature displays in the temperature button. The
actual temperature of the sapphire cooling window displays in the
temperature display next to the temperature button.
Temperature
pop-up window
Temperature
display
Temperature
button
Location of Temperature Button, Window and Display on
CoolView Treatment Screen