65
EXCEL V+ OPERATOR MANUAL
D2325, REV A, 2/19
Filling the Coolant Reservoir
The system requires clean distilled water. Systems are delivered without any water in them and must be
filled with distilled water upon installation. Occasionally, distilled water may need to be added to “top off”
the system reservoir.
To completely fill the coolant reservoir in the system:
1.
Turn off the system.
2.
Remove the reservoir fill cap at the rear of the system.
3.
Add water to the reservoir. Use only clean, unused distilled water provided. Carefully fill
the reservoir. Do not overfill.
4.
After the reservoir is full, turn the system keyswitch to the START position and release to
the ON position. This engages the water pump, allowing the coolant to be drawn into the
cooling system.
5.
Allow the system to run for 5 to 10 seconds.
6.
Turn the keyswitch to the OFF position.
7.
Top off the reservoir carefully.
8.
Repeat steps 4 thru 7 until the reservoir is full.
9.
Replace the fill cap.
CAUTION
Permanent damage to internal components will occur
if improper
coolant is used. One gallon of distilled water is shipped with every
system. Under no circumstances should ethylene glycol, tap water
or any liquid other than distilled water be used.
NOTE
The coolant capacity of the system is approximately one
half gallon.
CAUTION
Ensure that the fill cap has a small hole to vent the system. A similar
fill cap without a vent hole is used to transport the system.
Summary of Contents for Excel V+
Page 1: ...I Operator Manual...
Page 8: ...viii EXCEL V OPERATOR MANUAL D2325 REV A 2 19 THIS PAGE IS LEFT BLANK...
Page 14: ...6 EXCEL V OPERATOR MANUAL D2325 REV A 2 19 Remote Interlock Plug Footswitch...
Page 104: ...A 2 EXCEL V OPERATOR MANUAL D2325 REV A 2 19 THIS PAGE IS LEFT BLANK...
Page 110: ...C 2 EXCEL V OPERATOR MANUAL D2325 REV A 2 19 THIS PAGE IS LEFT BLANK...
Page 112: ...D 2 EXCEL V OPERATOR MANUAL D2325 REV A 2 19 THIS PAGE IS LEFT BLANK...