56
EXCEL V+ OPERATOR MANUAL
D2325, REV A, 2/19
“Connect Footswitch” advisory pop-up screen appears on the touch screen
with its corresponding error code(s).
“Water Low” advisory pop-up screen appears on the touch screen.
Error Code Guide
The excel V+ system utilizes state-of-the-art, self-diagnostic software. When the system detects an
abnormality during self-test or while the system is in READY mode, a pop-up screen will appear with a
corresponding error code(s). Some errors serve as a warning and are clearable by simply pressing the
button in the pop-up screen. Others, such as “connect footswitch” and “water low”, are accompanied
by an animated illustration showing what action to take. If the error or abnormal condition is successfully
cleared, the pop-up screen can be cleared by pressing the
button.
Probable Cause
The footswitch is not properly connected to the console.
Suggestion
Ensure that the footswitch is properly connected to the con-
sole.
NOTE
Press the
button to clear the advisory message.
Probable Cause
The system is low on water.
Suggestion
Turn off the system, and add water as instructed in the
Rou-
tine User Maintenance
topic. Restart the system to continue
with treatments.
NOTE
If an error cannot be cleared, or if an error persists or reappears,
note the error code(s) and contact Cutera Service (
refer to Appendix
D
).
Summary of Contents for Excel V+
Page 1: ...I Operator Manual...
Page 8: ...viii EXCEL V OPERATOR MANUAL D2325 REV A 2 19 THIS PAGE IS LEFT BLANK...
Page 14: ...6 EXCEL V OPERATOR MANUAL D2325 REV A 2 19 Remote Interlock Plug Footswitch...
Page 104: ...A 2 EXCEL V OPERATOR MANUAL D2325 REV A 2 19 THIS PAGE IS LEFT BLANK...
Page 110: ...C 2 EXCEL V OPERATOR MANUAL D2325 REV A 2 19 THIS PAGE IS LEFT BLANK...
Page 112: ...D 2 EXCEL V OPERATOR MANUAL D2325 REV A 2 19 THIS PAGE IS LEFT BLANK...