CU-ER1 Operator’s Manual _ver 3.00
Maintenance Activities
Frequency
Activity
Actions to be Taken
Daily
Check the Paramedic CU-ER1
for any error messages that
might have been generated
during the Daily Periodic Self-
Test.
If a message that the battery is low
is displayed, recharge the battery.
For any other error messages,
please call the manufacturer or its
designated service center.
Check supplies, accessories,
and spare parts for damage
and expiration.
If any supplies have expired,
replace them immediately.
Initiate complete self-test by
doing a Manual Self-Test.
If the self test detects any problem,
see the chapter on Troubleshooting.
Check the case of the
Paramedic CU-ER1
and the
accessories for any sign of
apparent damage.
If there is any apparent damage to
the case of the device, consult the
manufacturer.
Monthly and
after each use
Check for dirt contamination.
If there is dirt contamination, clean
the case as suggested in the section
on Cleaning.
After each use.
Ensure that the connector of
the defibrillator pad assembly
is disconnected from the ECG-
DEFIB port of the
Paramedic
CU-ER1
.
Disconnect the pad assembly from
the ECG-DEFIB port
73