CU-ER1 Operator’s Manual _ver 3.00
ECG Monitoring Mode
Input
Single channel ECG (Lead II) may be viewed on LCD and
printed.
Lead II ECG is acquired using a 3-wire ECG cable and
connector assembly
Lead Fault
“ATTACH PADS” prompt and dashed line on BPM value are
displayed when the ECG cable and connector assembly is
disconnected from the ECG-DEFIB port.
Heart Rate Display
Digital readout display from 30 to 300 bpm with an accuracy
of ±2 %
ECG Cable Length
3 meters
Common mode rejection Greater than 90 dB measured per AAMI EC 13
ECG size
•
5 mm/mV
•
10 mm/mV
•
20 mm/mV
•
AUTO: 0.3 to 1 mV signals are displayed with 10mm/mV
gain, outside of that range, the peak to peak value is
displayed as 10 mm on the LCD display.
Frequency Response
ECG Monitoring Mode: 0.3 to 40 Hz (-3 dB)
AED Mode (EMS): 1 Hz to 30 Hz
Patient Isolation
AED Mode
BF
ECG Monitoring Mode
BF
Event Storage
Internal Flash Memory
12 hours of event and ECG recording
SmartMedia Card
(32MB):
42 hours of event and ECG recording or 1 hour if voice
recording is enabled
ECG Record Printing
•
May be printed directly from the Paramedic CU-ER1
using a portable thermal printer.
•
May also be printed using a personal computer by
downloading the data to the PC using the data
management software from CU Medical Systems, Inc.
(CU Expert)
External Links
UART Port / IrDA Port
105