CU-ER1 Operator’s Manual _ver 3.00
Internal Battery Charging Indicator
The Paramedic CU-ER1 indicates that charging is in progress through the Battery Indicator
lamp. When charging is in progress, the Battery Indicator lamp is red and flashing. When
charging is finished, the Battery Indicator lamp is green and non-flashing.
Do not remove or change the internal battery pack. If the battery pack
reaches the end of its lifecycle, have it changed and disposed of by the
manufacturer or any of its authorized representatives.
The internal battery pack is not to be disposed of in fire nor incinerated.
It is not to be shorted nor disassembled.
It is not to be taken off its compartment and recharged with a charger other
than the built-in charger of the Paramedic CU-ER1.
If the internal battery pack leaks, call the manufacturer or any of its
authorized representatives immediately.
When the operating time of the internal battery pack becomes much shorter
than its initial operating time, have it replaced by the manufacturer or any of
its authorized representatives
The battery pack is not to be subjected to sharp impacts or concussions.
The battery pack is to be recharged within the temperature range 0° to 40°.
Recharging beyond this range may cause leakage of battery fluid or
heat generation. The performance and life of the battery pack may also
be impaired.
65