CU-ER1 Operator’s Manual _ver 3.00
Voice and Text Prompts, continued
Prompt Indication
BEGIN CPR, NOW
Voice and Text
You must administer cardiopulmonary resuscitation (CPR)
PUSH THE CHEST DOWN
FAST TWO INCHES
Voice and Text
Do compression by pushing the chest down fast two
inches.
GIVE TWO BREATHS.
Voice and Text
Indicates that you must give respiration to the patient.
BREATH. BREATH
Voice
Guides you with on the correct respiration delivery rhythm.
NO SHOCK ADVISED
Indicates that the patient has a non-shockable ECG
rhythm.
CHECK PULSE
Voice and Text
You must check the pulse of the patient.
IF NO PULSE, BEGIN CPR
Voice and Text
If there is no pulse, you must begin CPR
THE SHOCK BUTTON WAS
NOT PRESSED
Voice and Text
Indicates that a shockable rhythm has been detected and
a prompt to press the SHOCK button has been given but
the SHOCK button is not pressed within 15 seconds. The
device will discharge through its internal circuit.
ECG MONITORING MODE
Text
This indicates that the custom designed ECG MONITORING
CABLE AND CONNECTOR ASSEMBLY is attached to the
ECG-DEFIB port. In this mode, the Paramedic CU-ER1 can
do only acquisition and display of ECG. It does not analyze
the ECG for a shockable rhythm and it can not deliver a
defibrillating shock.
29