CU-ER1 Operator’s Manual _ver 3.00
5 Device Setup and Storage
5.1 Unpacking
•
Carefully inspect the packing container and the device for any damage that might
have been sustained during shipping.
•
Check the shipping list to ensure that the unit comes with the complete accessories.
It is important to have all the necessary accessories all the time. Make sure
that you have the complete accessories during unpacking.
If there are problems with the shipment, contact the distributor that shipped the product
to you.
5.2 Setup
The
Paramedic CU-ER1
is a self contained portable device. Please follow the following
setup procedures upon receiving your unit to ensure that the device is always ready for
any emergency.
5.2.1 Manual Self-Test
Conduct a Manual Self-Test to verify that the device is in good order. Initiate a Manual Self-
Test by performing the following steps:
a.
Press the UP and DOWN keypad buttons simultaneously.
b.
Press the ON/OFF button without releasing the UP and DOWN buttons.
c.
The
Paramedic CU-ER1
turns ON and displays the text prompt “SYSTEM CHECK”
d.
Follow the text prompts displayed by the device that tell you to press the Menu Keys
and the Shock Button one by one.
e.
Verify that the
Paramedic CU-ER1
displays the text prompt “SYSTEM OK”. The
Paramedic CU-ER1
shuts down automatically after the Manual Self-Test.
f.
If the
Paramedic CU-ER1
detects any fault in its system, it emits a beep and turns
the error indicator light ON momentarily (see section 9.1 – Self-Tests, for more
details).
g.
If a fault is detected, turn the
Paramedic CU-ER1
ON to see the error code.
h.
If a battery failure is detected, recharge the battery pack (see section 8.2 – Internal
Battery Pack Recharging, for details). For any other failure, contact CU Medical
Systems, Inc. or its authorized representative.
51