CU-ER1 Operator’s Manual _ver 3.00
You may scroll the highlighter between Compression-Breath ratio and ECG GAIN by
pressing the UP or DOWN key button.
You may change the Compression-Breath ratio or ECG GAIN setting by pressing the Menu
button while Compression-Breath ratio or ECG GAIN is highlighted, respectively. When the
parameter variable is highlighted, change its value by pressing the UP or DOWN button.
Compression Breath Ratio setting allows you to change the chest compression to
respiration ratio setting.
Two settings are available for Compression Breath Ratio:
•
30 : 2 -> CPR guide allocates time and beat for 30 chest compressions and 2
respirations per CPR cycle.
•
15 : 2 -> CPR guide allocates time and beat for 15 chest compressions and 2
respirations per CPR cycle.
ECG GAIN allows you to set the gain of the ECG display.
The possible ECG GAIN values are:
a. 5 mm/mV
b. 10 mm/mV
c. 20 mm/mV
d. AUTO
AUTO displays the ECG autoscaled (0.3 to 1 mV signals are displayed with 10mm/mV gain,
outside of that range, the peak to peak value is displayed as 10 mm on the LCD display).
47