Impedo
DUO
User Guide
53
5.1.2
Storer
Storer is the PQ recording subsystem. If it is running the two meter point status icons
are shown at the top of the OLED status display (see Section 4.1). See the Osprey Lite
user manual for more information on the restart procedure of the Storer.
5.2
Communicating with the Impedo Duo
All communication with the Impedo Duo is IP based. To secure communication over
open networks, all traffic is encrypted. The encryption requires two passwords – the
fleet password and the instrument key.
The fleet password is a shared secret across all servers and instruments belonging to
the fleet. The fleet password must be entered on the Impedo Duo using
Registration
(see Section 5.6).
The instrument key is unique to each instrument and comes preloaded on the
instrument. The key is also provided on the USB storage device shipped with the
instrument. In order to communicate with the instrument the instrument key must
be loaded into the server (see
Osprey Lite User Guide
for instructions on registering
an instrument with Osprey Lite).
The Impedo Duo communicates using several well-known ports. Ensure that ports
27030 – 27045 are not blocked by any firewall between the instrument and the
Osprey server. These ports must also be forwarded should any network address
translation occur between the instrument and the Osprey server.
Both the fleet password and the instrument key are required
before communication with the Impedo Duo is possible
The Impedo Duo communicate on ports 27030 – 27045
Summary of Contents for Impedo DUO
Page 85: ...Impedo DUO User Guide 85 Figure 49 Export harmonic selection...
Page 109: ...Impedo DUO User Guide 109 Appendices 8 Impedo DUO dimensions...
Page 110: ...Impedo DUO User Guide 110 8 1 Panel mounting plates 8 2 19 installation plate...
Page 111: ...Impedo DUO User Guide 111 9 Impedo DUO wiring diagram...
Page 113: ...Impedo DUO User Guide 113 11 Portable ImpedoDUO...
Page 117: ...Impedo DUO User Guide 117 Screw driver USB with meter key Ethernet cable...