Operation / Technical Manual
NORM-O-TEMP
®
, Model 111W
Page 42 of 82
5-1.2.
Internal Cleaning and Disinfecton Using Gigasept FF / Dry Storage
Procedure
For Circulating Water Units Utilizing CE Approved Gigasept FF (Product of
Schulke & Mayr) (Contact your local Cleaning Supplier). Gigasept FF Is Not
Available In the United States.
The following procedure disinfects the fluid circuit in these products. The
required tools/supplies are appropriate drain hose(s) and/or hose connector(s),
Gigasept FF, distilled water, and the appropriate AC electrical power
Procedure:
A.
Drain the water from the reservoir as described in Section 5-1.4.
B.
Turn unit off (“O” position).
C.
Fill the reservoir with the appropriate amount of Gigasept FF per the below
chart.
D.
Fill the reservoir with distilled water.
E.
Turn the unit on and circulate per the Chemical Cleaning Circulation Chart
(above).
F.
Drain the unit as instructed in Step A.
G.
Rinse the unit with Distilled Water three (3) times as described in Steps D-F. Be
sure to omit Gigasept FF.
H.
After the third rinse, drain the unit and fill the reservoir with distilled water and
circulate.
I.
Check the water with PH strips or other appropriate test method for Gigasept FF
detection. If Gigasept FF is detected, repeat Steps 4-6.
J.
If unit is to be
returned to service, replenish reservoir(s) with the appropriate
volume of distilled water.
K.
If
unit is being
placed in dry storage, continue with procedure.
L.
When all fluid has been removed from the unit, disconnect the drain hose(s) and
wipe unit clean.
M.
Unit is now ready for storage.
N.
Document Unit Maintenance.
UNIT
Gigasept FF
WATER
NORM-O-TEMP
®
, 111W
168 Milliliters 5.4 Liters