NORM-O-TEMP, Model 111W
The NORM-O-TEMP
®
, Model 111W hyperthermia System is used to keep a patient
comfortable by maintaining blanket/pad water temperature through conductive heat
transfer. The NORM-O-TEMP
®
hyperthermia system is composed of a heater,
circulating pump, safety high limit, and microprocessor board.
Distilled water is heated and pumped from the unit to a blanket/pad. The blanket/pad
rests under, around, and/or on top of the patient. The water circulates through the
blanket/pad and returns to the unit. When warmed water is circulated through the
blanket/pad, the patient is warmed. The unit is designed to operate based on the
temperature of the circulating water.
The recommended blankets/pads for use are listed in Figure 16 Accessories.
1-3.
Physical Description of the Norm-O-Temp
®
System
See Section 2 for specifications and certifications of the NORM-O-TEMP
®
system.
1-3.1.
External Features and Descriptions – Front View
The external features in Figure 1 of the NORM-O-TEMP
®
unit are described as
follows:
A.
The control panel is composed of pressure sensitive touch switches and an LED
display. An expanded description of the membrane control panel is presented in
Section 1-3.5.
B.
The operating instructions printed below the control panel describe the
necessary steps to operate the unit.
C.
The power switch is a bevel rocker switch labeled “I” (on) at the top and “O”
(off) at the bottom.
D.
The four rubber feet help the unit to mount on the I.V. stand (Catalog #118), the
low profile stand (Catalog #119) or allow the unit to sit on any flat surface.
E.
The water fill opening is where the operator pours distilled water to fill the
reservoir.
F.
The red, Power Fail LED is the visual indicator when power is disconnected with
the switch still in its on position or when the independent mechanical (High
Limit) safety is activated.