Operation / Technical Manual
NORM-O-TEMP
®
, Model 111W
Page 27 of 82
Electrical System
Electrical Characteristics:
230V, 50Hz, 4.8 Amps
240V, 50Hz, 4.6 Amps
Heater Power:
800 watts
Power Cord:
16/3 Detachable, IEC 320
Leakage Current:
Under 500
µ
a 230/240 VAC
Fuse
5 Amp in IEC 320 connector
Mains Supply Isolation:
Two-Pole Mains Switch
Warranty
1 year parts. Parts and labor if returned to
factory.
Additional Warranty Available
Safety System
Maximum High Control Setting
42°C
Primary Temperature Limit:
43.5°C ± 0.6°C
Secondary Temperature Limit:
44.5°C ± 0.6°C
Independent Mechanical Temperature
Limit:
46°C ± 0.6°C
Environmental Conditions
(during storage and transportation)
Ambient Temperature (transportation
and storage):
-40°C –
+
50°C
Humidity (transportation and storage):
20% – 95%
UL Electrical Classification
Equipment is Class I.
Equipment is Type BF.
CERTIFICATIONS
Figure 6. NORM-O-TEMP
®
, Model 111W Features
Medical Electrical Equipment
In Accordance With:
UL 60601-1 2
nd
Ed.
IEC 60601-1-2 3
rd
Ed.
ASTM F-2196-2