Figure 3-4. Port Flushing Kit and Instruments Connections
3.
By-Pass Safety Valve:
Prevents excessive water pressure from
damaging delicate instruments.
Ten Position (Ports) Terminal Block:
Permits up to Ten
instruments to be cleaned at once.
Terminal Block Retainer:
Designed to securely hold Ten
Position Terminal Block (see).
Instrument Support:
Supplied in Instrument Tray to hold
instruments in position during cleaning cycle.
4.
5.
6.
3.4.2 Operation
Prepare and process lumened surgical instruments as follows:
1. Insert Ten Position Terminal Block (see Figure 3-4) into Terminal
Block Retainer. Slide retainer over Instrument Tray rear wall.
NOTE: Ten Position Terminal Block may be repositioned for
convenience by sliding Terminal Block Retainer left or right.
Attach Instrument Support to Instrument Tray front wall with tips
or ends pointing down for drainage.
Connect Instrument Tubing to Terminal Block using quick-
disconnect side of tubes.
Attach large main tube to Terminal Block and to Large Quick-
Connect Fitting on underside of lid.
Position instruments as follows:
a. Position up to Ten instruments into Instrument Tray and
connect Luer-Lock of Instrument Tubing to instrument luer-
lock connecting port.
2.
3.
4.
5.
Crest Healthcare Console Junior R:1
Operating Instructions
3-9
Operator Manual
Large Quick-Connect Fitting
Terminal Block Retainer
Terminal Block
Instrument Support
Lumen Instrument Instrument Tubing
Instrument Tray (Included With System)