![Crest Audio Console Junior Operator'S Manual Download Page 12](http://html1.mh-extra.com/html/crest-audio/console-junior/console-junior_operators-manual_2683278012.webp)
2.3 Technical
Specifications
The following specifications are intended to describe the technical
information located on your unit nameplate and to state other relevant
information. Check the previously supplied equipment drawing and/or
identification nameplate for proper voltage and amperage.
NOTE: Electrical and plumbing services must be flexible or have a
quick disconnect so Ultrasonic Cleaning System may be moved
away from the wall for servicing.
2.3.1 Overall Size (W x L x H)
Sonic Cleaner - 35 x 30 or 36 x 38-1/2"
(889 x 762 or 914 x 978 mm)
•
2.3.2 Electrical
Power Supply:
240 V, 60 Hz, 1-Phase (3 wire) or 208 V, 3-Phase
(5 wires including neutral and ground),208 V, 1-Phase (4 wires).
Power Cord Specification:
• Sonic Cleaner - Minimum 8 AWG, rated 194°F (90°C), 600 V
NOTE: Power cord and wall receptacle are not furnished by
Crest. Verify power cord and wall receptacle used meet all
national and local electrical codes (ensure UL817 compliant).
Only use green/yellow conductors for connection to
protective conductor terminals.
Disconnect Switch:
• Sonic Cleaner - 60 Amp rated, lockable, fusible, UL Listed E2875
and 154828 (or equivalent)
2.3.3 Other Facility
Requirements:
Drain:
1-1/4" CPVC Male Slip (trap and waste lines are not provided
by Crest).
Distilled (Purified) Water (Optional):
1/2" CPVC Male Slip (15-
50 psi); 100-120°F (38-49°C).
Hot Water:
1/2" CPVC Male Slip (15-50 psi); 100-120°F (38-49°C).
Cold Water:
1/2" CPVC Male Slip (15-50 psi); 40-60°F (4-16°C).
NOTE: Customer is responsible for compliance with applicable
local and national codes and regulations.
2.3.4 Weight
• Sonic Cleaner - 15-gal: 452 lb (205 kg)
20-gal: 545 lb (247 kg)
2.3.5 Seismic Anchorage
A Seismic Anchorage System is available for high-risk seismic zones.
Crest Healthcare Console Junior R:1
2-3
Important User Information
Operator Manual