00 | Page 61
www.covvi.com | Unit 4, Quayside Business Park, Leeds, LS10 1DJ | 020 3949 9500
13.4 Input Settings
Clicking on ‘Input Settings’ takes you to a graph where you can see your patient’s inputs in real time and
change the thresholds to map the speed of the hand to their signals. This allows your patient to maximise
performance of their Nexus Hand. There is the option of one input reading or two input readings. If the user
selects single site in operating modes, the signal is shown on Input A. If the user selects dual site, the signals
are shown on Input A and B.
Example Input
Graph
13.0 COVVI Go App
On Threshold -
This shows the signal strength needed for the hand to move.
Max Threshold -
This shows the signal strength at which the hand will move at full speed. Any signals above
this level will make no difference to speed.
There are four primary thresholds:
1. Input A On Threshold
2. Input A Max Threshold
3. Input B On Threshold
4. Input B Max Threshold