00 | Page 39
www.covvi.com | Unit 4, Quayside Business Park, Leeds, LS10 1DJ | 020 3949 9500
11.0 The COVVI Power Supply
• The COVVI Power Supply should be fitted by a certified prosthetist.
• This product uses semiconductors that can be damaged by electrostatic discharge (ESD).
• This product is not designed to be disassembled or serviced. Damage due to inappropriate handling is
not covered by the warranty.
• COVVI has the right to void the warranty of all products that have any type of modifications or damage
caused by any unauthorised or untrained personnel.
• Individuals who are exposed to hazardous environments that contain flammable liquid or gas should NOT
operate the COVVI Power Supply.
• Please make sure the Power Supply is OFF before connecting/disconnecting the prosthetic device as this
will avoid any unpredictable spikes in the electric current from the batteries. Power on the Power Supply
once the prosthetic device is properly connected to the prosthesis.
• Do
NOT
expose the COVVI Power Supply to an open flame or submerge it in water.
• Do
NOT
attempt to use the prosthetic device while the batteries are charging. Once the batteries are
charging, the power automatically turns off. If for any reason the power does not turn off while charging,
using the prosthetic device can be potentially unsafe.
• Do
NOT
use the Power Supply if there is any visible sign of damage to the Power Supply charger, Power
Plug and/or Cables.
• Do
NOT
fit the batteries into a small, enclosed space. The batteries are designed to inflate when they fail.
Stopping this inflation may cause the generation of excessive heat, injury, or death.
• Ensure access to wall plug to enable easy isolation if required.
• The COVVI Power Supply should only be used with COVVI batteries.
11.2 Warnings
Summary of Contents for AK140E316
Page 2: ...Anything Is Possible...