SurgiStat Low-Power Electrosurgical Generator Service Manual
Understanding Error Codes and Audio Tones
E7
Temperature error
Internal temperature of the
generator exceeded limit.
1) Turn the generator off.
2) Allow the generator to cool
for 20 minutes.
3) Restart the generator.
4) If the error code reappears,
record the number and
contact your Covidien Service
Center.
E8
Connector sense error
The generator shall monitor
the connection of the main
cable between the main and
display boards.
1) If the cable is disconnected, an
E8 error will be displayed.
2) The generator cannot be
activated while the error
condition is present.
3) Restart the generator to
remove the error condition.
Error Code
Description
Recommended Action