Functional (Operational) Checks
Generator Operation
Activating the Generator
Activate the generator by pressing the appropriate button on the handswitch or
footswitch.
To avoid the possibility of an electrosurgical burn to either the patient or the physicians, do not
allow the patient to come in contact with a grounded metal object during activation. When
activating the unit, do not allow direct skin contact between the patient and the physician.
To avoid the possibility of a burn to the patient, when using a split pad do not activate the unit if
the solid pad indicator is illuminated green or the red alarm indicator remains illuminated red. This
could indicate improper pad placement or a faulty NEM contact-monitoring circuit.
Remove any jewelry from the patient before activation.
Examine all accessories and connections to the electrosurgical generator before use. Ensure that
the accessories function as intended. Improper connection may result in arcs, sparks, accessory
malfunction, or unintended surgical effects.
When not using active accessories, place them in a holster or in a clean, dry, non-conductive, and
highly visible area not in contact with the patient. Inadvertent contact with the patient may result
in burns.
Studies have shown that smoke generated during electrosurgical procedures can be potentially
harmful to patients and the surgical team. These studies recommend adequately ventilating the
smoke by using a surgical smoke evacuator or other means.
1
1. U.S. Department of Health and Human Services. National Institute for Occupational Safety and
Health (NIOSH). Control of Smoke from Laser/Electric Surgical Procedures. HAZARD CONTROLS,
Publication No. 96-128, September, 1996.
Monopolar Cut
Select the desired mode of operation (Pure Cut or Blend), then select the
power settings by rotating the Cut and Blend Power Control Dial.
Monopolar Coag
Select the coag mode of operation (Desiccate or Fulgurate), then select
the power setting by rotating the Coag and Bipolar Power Control Dial.
Bipolar
Select the mode of operation for Bipolar (coagulation or fulguration),
then select the bipolar power settings by rotating the Coag and Bipolar
Power Control Dial.
Notice
One footswitch can activate either monopolar or bipolar operation.
Precaution
Summary of Contents for SurgiStat
Page 1: ...Service Manual SurgiStat TM Low Power Electrosurgical Generator ...
Page 2: ......
Page 3: ...Service Manual SurgiStat TM Low Power Electrosurgical Generator 1079008 ...
Page 10: ......
Page 34: ......
Page 54: ......
Page 74: ......
Page 88: ......
Page 110: ......
Page 111: ......