SurgiStat Low-Power Electrosurgical Generator Service Manual
EMC Compliance
Voltage dips,
short
interruptions
and voltage
variations on
power supply
input lines EC
61000-4-11
<5%
U
t
(>95%
dip in
U
t
) for 0.5
cycle
40%
U
t
(60% dip
in
U
t
) for 5 cycles
70%
U
t
(30% dip
in
U
t
) for 25 cycles
<5%
U
t
(>95%
dip in
U
t
) for 5
sec.
<5%
U
t
(>95%
dip in
U
t
) for 0.5
cycle
40%
U
t
(60% dip
in
U
t
) for 5 cycles
70%
U
t
(30% dip
in
U
t
) for 25 cycles
<5%
U
t
(>95%
dip in
U
t
) for 5
sec.
Mains power quality should be
that of a typical commercial or
hospital environment. If the user
of the SurgiStat requires
continued operation during
power mains interruptions, it is
recommended that the SurgiStat
be powered from an
uninterruptible power supply ora
battery.
Power frequency
(50/60Hz)
magnetic field
IEC 61000-4-8
3A/m
3 A/m
Power frequency magnetic fields
should be at levels characteristic
of a typical location in a typical
commercial or hospital
environment.
Note:
U
t
is the a.c. mains voltage prior to application of the test level.
Guidance and manufacturer’s declaration – electromagnetic immunity
Summary of Contents for SurgiStat
Page 1: ...Service Manual SurgiStat TM Low Power Electrosurgical Generator ...
Page 2: ......
Page 3: ...Service Manual SurgiStat TM Low Power Electrosurgical Generator 1079008 ...
Page 10: ......
Page 34: ......
Page 54: ......
Page 74: ......
Page 88: ......
Page 110: ......
Page 111: ......