5-4
SurgiStat Low-Power Electrosurgical Generator Service Manual
Installation and Placement
Installation and Placement
Place the SurgiStat electrosurgical generator on any flat surface with a tilt angle of not
more than 10 degrees. The generator relies on natural convection cooling. Do not block
the rear vents.
Carts with conductive wheels are recommended. For details, refer to the procedures for
your institution or to local codes.
Ensure that air flows freely on all sides of the generator. Provide at least 10 –15 cm (4 – 6
in.) of space from the sides and top of the generator for cooling.
Normally, the top, sides, and rear panel are warm when you use the generator
continuously for extended periods of time.
Functional (Operational) Checks
Upon initial installation of the generator, perform the following checks. Refer to the
figures in Section 2, Controls, Indicators, and Receptacles, for the location of the controls
and connectors.
How to Set Up and Start the SurgiStat
1. Verify that the power switch is in the Off (O) position and that no accessories are
connected to the generator.
2. Connect a hospital-grade power cable to the AC power cable receptacle on the back
of the generator, then to a properly grounded wall outlet.
3. Connect a two-pedal footswitch to the matching footswitch receptacle on the back of
the generator. Use only a Covidien footswitch with the SurgiStat generator. Although
other types of footswitches may fit, they may not be compatible with this
electrosurgical generator. Use of an incompatible footswitch may cause unexpected
output.
4. Turn on the generator by switching the power switch to the On (|) position.
Warning
Connect the power cord to a properly polarized and grounded power source with the frequency
and voltage characteristics that match those listed on the back of the generator.
Precaution
At no time should you touch the active electrode or bipolar forceps. A serious burn could result.
Important
Do not connect a patient return electrode to the front of the generator at this time.
Summary of Contents for SurgiStat
Page 1: ...Service Manual SurgiStat TM Low Power Electrosurgical Generator ...
Page 2: ......
Page 3: ...Service Manual SurgiStat TM Low Power Electrosurgical Generator 1079008 ...
Page 10: ......
Page 34: ......
Page 54: ......
Page 74: ......
Page 88: ......
Page 110: ......
Page 111: ......