SurgiStat Low-Power Electrosurgical Generator Service Manual
3-9
EMC Compliance
Guidance and manufacturer’s declaration – electromagnetic emissions
The SurgiStat Low-Power Electrosurgical Generator is intended for use in the electromagnetic
environment listed below. The customer or the user of the SurgiStat Generator should assure that
it is used in such an environment.
Emissions test
Compliance
Electromagnetic environment -
guidance
RF Emissions CISPR 11
Group 2
The SurgiStat must emit
electromagnetic energy in order to
perform its intended function.Nearby
electronic equipment maybe affected.
RF Emissions CISPR 11
Class A
The SurgiStat is suitable for use in all
establishments other than domestic
and those directly connected to the
public low-voltage power supply
network that supplies buildings used in
domestic purposes.
Harmonic emissions
IEC 61000-3-2
Class A
Voltage fluctuations/
flicker emissions
IEC 61000-3-3
Complies
Guidance and manufacturer’s declaration – electromagnetic immunity
The SurgiStat is intended for use in the electromagnetic environment listed below. The customer
or the user of the SurgiStat should assure that it is used in such an environment.
Immunity test
IEC 60601 test
level
Compliance
leve
l
Electromagnetic environment -
guidance
Electrostatic
discharge (ESD)
IEC 61000-4-2
±6 kV contact
±8 kV air
±6 kV contact
±8 kV air
Floors should be wood, concrete
or ceramic tile. If floors are
covered with synthetic material,
the relative humidity should be at
least 30%.
Electrical fast
transient/burst
IEC 61000-4-4
±2 kV for power
supply lines
±1 kV for input/
output lines
±2 kV for power
supply lines
±1 kV for input/
output lines
Mains power quality should be
that of a typical commercial or
hospital environment.
Surge IEC
61000-4-5
±1 kV differential
mode
±2 kV common
mode
±1 kV differential
mode
±2 kV common
mode
Mains power quality should be
that of a typical commercial or
hospital environment.
Summary of Contents for SurgiStat
Page 1: ...Service Manual SurgiStat TM Low Power Electrosurgical Generator ...
Page 2: ......
Page 3: ...Service Manual SurgiStat TM Low Power Electrosurgical Generator 1079008 ...
Page 10: ......
Page 34: ......
Page 54: ......
Page 74: ......
Page 88: ......
Page 110: ......
Page 111: ......