6-2
SurgiStat Low-Power Electrosurgical Generator Service Manual
Cleaning
Covidien recommends that periodic inspection and performance testing be performed
every six months. A qualified biomedical technician should conduct this testing to ensure
that the unit is operating effectively and safely.
Cleaning
After each use, clean the generator.
1. Turn off the generator, and unplug the power cord from the wall outlet.
2. Thoroughly wipe all surfaces of the generator and power cord with a mild cleaning
solution or disinfectant and a damp cloth. Follow the procedures approved by your
institution or use a validated infection control procedure.
Periodic Inspection
Every six months, visually inspect the generator for signs of wear or damage. In particular,
look for any of the following problems:
•
Damage to the power cord
•
Damage to the power cable receptacle
•
Obvious damage to the unit
•
Damage to any receptacle
•
Accumulation of lint or debris in or around the unit
Warning
Electric Shock Hazard
Always turn off and unplug the generator before cleaning.
Precaution
Do not allow fluids to enter the generator chassis.
Do not sterilize the generator.
Notice
Do not clean the generator with abrasive cleaning or disinfectant compounds, solvents, or other
materials that could scratch the panels or damage the generator.
Summary of Contents for SurgiStat
Page 1: ...Service Manual SurgiStat TM Low Power Electrosurgical Generator ...
Page 2: ......
Page 3: ...Service Manual SurgiStat TM Low Power Electrosurgical Generator 1079008 ...
Page 10: ......
Page 34: ......
Page 54: ......
Page 74: ......
Page 88: ......
Page 110: ......
Page 111: ......