SurgiStat Low-Power Electrosurgical Generator Service Manual
7-5
Understanding Error Codes and Audio Tones
Error Code
Description
Recommended Action
F1
on the Cut/Blend
display
F1
on the Desiccate/
Fulgurate/Bipolar display
Handswitch or monopolar
footswitch cut pedal may be
stuck
Handswitch or monopolar
footswitch coag pedal may
be stuck
.
1) Turn off, then turn on the
generator. Do not press
buttons or accessory
activation devices during the
self-test.
2) If the error coder reappears,
disconnect all accessories.
Turn off, then turn on the
generator again.
3) If the problem persists, replace
the handpiece or footswitch
and repeat the restart.
4) If the error code reappears,
record the number and
contact your Covidien Service
Center.
F2
Simultaneous activation
error
Cut and Coag buttons
activated simultaneously
(pencil or footswitch).
The generator does not allow
simultaneous activation of the Cut
and Desiccation modes. Release
either the Cut or Coag button on
the handpiece, or the Cut or Coag
pedal on the footswitch.
F3
Foot activation error
Footswitch Cut or Coag
pedal pressed while in
Bipolar Foot Control and the
generator is in Bipolar mode.
Release pressure on the handpiece
buttons. Use the footswitch to
activate bipolar, or switch to
monopolar mode.
E4
High voltage error
Line voltage error (Line
voltage too high.
1) Turn the generator off.
2) Verify that the generator is
connected to the correct line
voltage.
3) If the error code reappears,
record the number and
contact your Covidien Service
Center.
Summary of Contents for SurgiStat
Page 1: ...Service Manual SurgiStat TM Low Power Electrosurgical Generator ...
Page 2: ......
Page 3: ...Service Manual SurgiStat TM Low Power Electrosurgical Generator 1079008 ...
Page 10: ......
Page 34: ......
Page 54: ......
Page 74: ......
Page 88: ......
Page 110: ......
Page 111: ......