Operating Procedures
5-4
User’s Manual
5.2.2 USB Menu
To access the USB menu when a USB memory device is connected:
Press the
MENU
key several times, until the USB Menu appears:
Figure 5-4.
Selecting the USB Menu
In case of high priority alarm activation the ventilator will automatically display the alarm page. To return to
the USB Menu, press the MENU
key.
The adjustable parameters in this menu are:
•
Transfer continuously
•
Transfer trends
•
Erase key
5.2.3 Transfer Continuously
Up to 48 hours worth of data can be transferred from a ventilator to a USB memory device.
To record continuously, the USB memory device must be permanently connected to the ventilator and ventila-
tion is active.
The following data will be recorded to the USB memory device:
•
Monitoring: pressure, inspired flow, exhaled flow and leak waveforms.
•
Trends: leaks, VTI, VTE, Rate, I:E, M. Vol, PIP and PEEP measurements.
The data can be accessed by a doctor or service provider using the Puritan Bennett™ Respiratory Insight Soft-
ware.
Figure 5-5.
Selecting Transfer Continuously
To transfer continuous data from a ventilator to a USB memory device:
1. Use the
UP
or
DOWN
arrow keys to place the cursor at the “Transfer
Continuously” position.
2. Press
ENTER
.
Summary of Contents for Puritan Bennett 560
Page 1: ...User s Manual Puritan Bennett 560 Ventilator TM...
Page 6: ...This page intentionally blank iv User s Manual...
Page 10: ...viii User s Manual This page is intentionally blank...
Page 14: ...4 User s Manual This page is intentionally blank...
Page 32: ...Safety Information 1 18 User s Manual Figure 1 4 Location of Labels Bottom View 7 13 1 4 3...
Page 98: ...5 10 User s Manual This page is intentionally blank...
Page 128: ...C 2 User s Manual This page is intentionally blank...
Page 134: ...E 4 User s Manual This page is intentionally blank...
Page 148: ...Index Index 6 User s Manual This page intentionally blank...
Page 149: ......