Alarms and Troubleshooting
User’s Manual
3-7
3.6
Re-activating Alarms
Alarms that have been paused and whose activation conditions continue to exist can be reactivated.
To reactivate alarms, proceed as follows:
1. Press the
MENU
key to access the Alarm Setting menu, if this is not the menu currently
displayed.
2. Press the
DOWN
key to position the cursor on the “Alarm Logs” line, if this is not
already
the case. Refer to the following graphic:
Figure 3-6.
Reactivating Alarms
3. Press the
ENTER
key, to confirm access to the “Alarm Logs” menu.
4. Press the
UP
key to position the cursor on the “USER’S CLEAR ALERTS” line. Refer to the
following graphic:
Figure 3-7.
Alarm Logs
5. Press the
ENTER
key for at least three (3) seconds. The following events occur:
•
A “beep” sounds.
•
An audible alarm sounds.
•
An alarm indicator illuminates.
•
The messages of all active alarms are displayed in a loop in the Ventilation and Alarm
menus.
•
The Audio Paused symbol
disappears (if it was displayed).
•
The Alarm Paused symbol
disappears.
Summary of Contents for Puritan Bennett 560
Page 1: ...User s Manual Puritan Bennett 560 Ventilator TM...
Page 6: ...This page intentionally blank iv User s Manual...
Page 10: ...viii User s Manual This page is intentionally blank...
Page 14: ...4 User s Manual This page is intentionally blank...
Page 32: ...Safety Information 1 18 User s Manual Figure 1 4 Location of Labels Bottom View 7 13 1 4 3...
Page 98: ...5 10 User s Manual This page is intentionally blank...
Page 128: ...C 2 User s Manual This page is intentionally blank...
Page 134: ...E 4 User s Manual This page is intentionally blank...
Page 148: ...Index Index 6 User s Manual This page intentionally blank...
Page 149: ......