Specifications
A-8
User’s Manual
A.8
USB
A.9
Pneumatic
Table A-13.
USB Memory Device Specifications
Characteristics
Supported formats
USB compatibility
USB flash memory USB 2.0 or USB 1.1
Memory file format
USB 32 bit format (sector size: 512 - 2,048 bytes)
Number of files
Maximum 999
USB size
128 MB to 4 GB
Table A-14.
Data Transfer Characteristics
Ventilator data description
Capacity
Trends capacity
86 MB
Events capacity
512 KB or 5,500 events
Monitorings capacity
42 MB / 48 hours
Table A-15.
Airway Resistances
Inspiratory
Exhalation
1.0 mbar at 30 lpm flow ± 0.1 mbar
0.5 mbar at 30 lpm ± 0.1 mbar
3.7 mbar at 60 lpm flow ± 0.1 mbar
1.1 mbar at 60 lpm ± 0.1 mbar
Table A-16.
Patient Circuit Resistances
a
a. Includes exhalation valve
Adult Double Limb
Pediatric Double Limb
2 mbar at 60 lpm flow
b
b. Values obtained from the manufacturer’s directions for use.
2 mbar at 30 lpm flow
Table A-17.
Air Inlet Resistance (Filter)
1.1 cmH
2
O (1.079 mbar) at 30 lpm flow
Table A-18.
Oxygen Inlet Specifications
Maximum
pressure
Maximum
flow
50 kPa (7 psi)
15 lpm
Table A-19.
Performance Specifications
Working
pressure
Sound pressure
level
Maximum
pressure limit
Internal
compliance
(ventilator)
Inspiratory
triggering
response time
(Ttr)
5 mbar –
55 mbar
30 dBA (per NF
EN ISO 17510-1
test conditions)
60 mbar
.0001 l/mbar
100
ms
Summary of Contents for Puritan Bennett 560
Page 1: ...User s Manual Puritan Bennett 560 Ventilator TM...
Page 6: ...This page intentionally blank iv User s Manual...
Page 10: ...viii User s Manual This page is intentionally blank...
Page 14: ...4 User s Manual This page is intentionally blank...
Page 32: ...Safety Information 1 18 User s Manual Figure 1 4 Location of Labels Bottom View 7 13 1 4 3...
Page 98: ...5 10 User s Manual This page is intentionally blank...
Page 128: ...C 2 User s Manual This page is intentionally blank...
Page 134: ...E 4 User s Manual This page is intentionally blank...
Page 148: ...Index Index 6 User s Manual This page intentionally blank...
Page 149: ......