Specifications
User’s Manual
A-7
A.7
Environmental
The following environmental conditions shall be observed:
Under extreme conditions of use that are beyond the recommendations above but within the limits of a supply
voltage of –20%, compared to the nominal temperature or the combination of a temperature of 45 °C
(113 °F) and humidity of 75% RH, the ventilator should not malfunction nor endanger the user. However,
operating the device for prolonged periods or repeatedly under such extreme conditions could result in prema-
ture aging of components and more frequent maintenance.
Minimum Peak Inspiratory Pressure
(Min PIP)
Range: PIP- 20% (not adjustable in pressure breath)
Range: 2-52 in volume breath
Resolution: N/A
Maximum Peak Inspiratory Pressure
(Max PIP)
Range: PIP+ 20 % (not adjustable in pressure breath)
Range: 12-60 in volume breath
Resolution: N/A
Minimum inspiratory time
(Min I time)
Range: 0.1 to 2.8 s
Resolution: 0.1 s
Default value: AUTO (Rise time + 300 ms)
Depends on: Max I Time, Backup R, Rise time
Maximum inspiratory time
(Max I time)
Range: 0.8 to 3 s
Resolution: 0.1 s
Default value: AUTO {Min [3 s; (30/R-Rate)]}
Depends on: Min I Time, R-Rate
Minimum Fraction of Inspired Oxygen
(Min FIO
2
)
Range: 18 to 90 %
Resolution: 1 %
Default value: OFF
Depends on: Max FIO
2
Maximum Fraction of Inspired Oxygen
(Max FIO
2
)
Range: 30 to 100 %
Resolution: 1 %
Default value: OFF
Depends on: Min FIO
2
Table A-10.
Ventilator Range, Resolution, and Accuracy (Continued)
Ventilator Settings
Range, Resolution, and Accuracy
Table A-11.
Environmental Conditions for Storage or Transport
Temperature
Humidity
Atmospheric
pressure
Altitude
-40 °C to +70 °C (-40 °F to
+158 °F)
10% to 95% RH
500 hPa to 1060 hPa
(7 psi to 15.4 psi)
-152 m to 3964 m
(500 ft to 13,000 ft)
Table A-12.
Environmental Conditions for Operation
Temperature
Humidity
Atmospheric
pressure
Altitude
+5 °C to 40 °C (+41 °F to
104 °F)
10% to 95% RH
600 hPa to 1100 hPa
(8.7 psi to 16.0 psi)
-152 m to 3964 m
(-500 ft to 13,000 ft)
Summary of Contents for Puritan Bennett 560
Page 1: ...User s Manual Puritan Bennett 560 Ventilator TM...
Page 6: ...This page intentionally blank iv User s Manual...
Page 10: ...viii User s Manual This page is intentionally blank...
Page 14: ...4 User s Manual This page is intentionally blank...
Page 32: ...Safety Information 1 18 User s Manual Figure 1 4 Location of Labels Bottom View 7 13 1 4 3...
Page 98: ...5 10 User s Manual This page is intentionally blank...
Page 128: ...C 2 User s Manual This page is intentionally blank...
Page 134: ...E 4 User s Manual This page is intentionally blank...
Page 148: ...Index Index 6 User s Manual This page intentionally blank...
Page 149: ......