Installation and Assembly
User’s Manual
4-9
.
Figure 4-8.
Double Limb Patient Circuit
Figure 4-9.
Close-up of Exhalation Valve Tube and Proximal Pressure Tube
Figure 4-9
shows details of the connections of the proximal pressure tube (
Figure 4-8
, item 6) and
the exhalation valve tube (
Figure 4-8
, item 8).
To connect a single limb circuit without an exhalation valve (NIV only):
refer to
Figure 4-10
.
1. Inspect the components of the patient circuit for any signs of damage, such as cracks (which
might cause leakage). Do not use damaged components to assemble the patient circuit.
2. Install the bacteria filter (item 1) on the TO PATIENT
outlet port, as shown.
3. Attach one end of the short circuit tubing (item 4) to the filter (item 1)
4. Attach the other end of the circuit tubing (item 4) to the inlet port of the humidifier (item 2).
Note:
Although shown here, the humidifier (item 2), water traps (item 3), and their connecting tubes are not
included with the patient circuit or ventilator. Contact your supplier for more information.
See also
Figure 4-9
on page 4-9.
1
2
11
4
10
9
8
6
5
7
3
13
12
14
Proximal Pressure
Tube
Exhalation
Valve Tube
Inhalation Port
Summary of Contents for Puritan Bennett 560
Page 1: ...User s Manual Puritan Bennett 560 Ventilator TM...
Page 6: ...This page intentionally blank iv User s Manual...
Page 10: ...viii User s Manual This page is intentionally blank...
Page 14: ...4 User s Manual This page is intentionally blank...
Page 32: ...Safety Information 1 18 User s Manual Figure 1 4 Location of Labels Bottom View 7 13 1 4 3...
Page 98: ...5 10 User s Manual This page is intentionally blank...
Page 128: ...C 2 User s Manual This page is intentionally blank...
Page 134: ...E 4 User s Manual This page is intentionally blank...
Page 148: ...Index Index 6 User s Manual This page intentionally blank...
Page 149: ......