Operating Procedures
User’s Manual
5-5
•
The cursor changes to the plus/minus symbol.
•
The parameter selected to be modified flashes.
3. Press
UP
or
DOWN
to change the selected parameter’s value.
4. Press
ENTER
to confirm the new parameter setting.
•
The new parameter setting is displayed continuously.
•
The cursor is placed at the
STOP
position.
5. To manually stop continuous transfer, press the
ENTER
key.
If a parameter change is not confirmed by pressing
ENTER
before seven (7) seconds elapse, the ventila-
tor resets the parameter to its previous value.
5.2.4 Transfer Trends
Up to one year’s worth of trend data can be transferred from a ventilator to a USB memory device.
Ventilation trends such as leaks, VTI, VTE, Rate, I:E, M. Vol, PIP and PEEP measurements can be transferred
from the ventilator to a USB memory device.
The data can be accessed by a doctor or service provider using the Puritan Bennett™ Respiratory Insight Soft-
ware.
Figure 5-6.
Selecting Transfer Trends
To transfer trend data from a ventilator to a USB memory device:
1. Use the
UP
or
DOWN
arrow keys to place the cursor at the “Transfer Trends”
position.
2. Press
ENTER
.
•
The cursor changes to the plus/minus symbol.
Note:
All ventilator menus remain accessible during transfer time.
The message “TRANSFER IN PROGRESS... REMAINING TIME” is displayed during the transfer time.
Other functions of the USB memory device are not available during continuous recording,
If the memory capacity on the USB memory device is insufficient the message “TRANSFER NOT ALLOWED -
USB CAPACITY INSUFFICIENT” is displayed and data transfer is not allowed. Delete the data on the USB
memory device before restarting data transfer.
In case of USB memory device disconnection or transfer error, the message “TRANSFER ERROR - USB
DISCONNECTION” or “TRANSFER ERROR - TECHNICAL PROBLEM” is displayed. In this case restart the
transfer process. If the problem persists contact your technical service.
Summary of Contents for Puritan Bennett 560
Page 1: ...User s Manual Puritan Bennett 560 Ventilator TM...
Page 6: ...This page intentionally blank iv User s Manual...
Page 10: ...viii User s Manual This page is intentionally blank...
Page 14: ...4 User s Manual This page is intentionally blank...
Page 32: ...Safety Information 1 18 User s Manual Figure 1 4 Location of Labels Bottom View 7 13 1 4 3...
Page 98: ...5 10 User s Manual This page is intentionally blank...
Page 128: ...C 2 User s Manual This page is intentionally blank...
Page 134: ...E 4 User s Manual This page is intentionally blank...
Page 148: ...Index Index 6 User s Manual This page intentionally blank...
Page 149: ......