Alarms and Troubleshooting
3-2
User’s Manual
3.1
Alarm Level of Priority
The alarm hierarchy for signalling the level of alarm criticality is listed below.
•
Very High Priority (VHP): Immediate critical situation; ventilation is impossible:
Continuous Sound Signaling / With or Without Continuous Red LED Illumination / With or
Without Message / With or Without Display Lighting (it is possible for an alarm condition to
occur that may not have
both
a message and lighting).
•
High Priority (HP): Critical situation in the short term; ventilation is potentially
compromised:
High Speed Intermittent Sound Signaling / Flashing Red LED Illumination / With Message /
With Display Lighting
•
Medium Priority (MP): Critical situation in the long term; ventilation is not affected
in the short term:
Medium Speed Intermittent Sound Signaling / Flashing Yellow LED Illumination / With
Message / With Display Lighting
Note:
If there is no corrective action and if the audible alarm is not inhibited (Audio Paused) or reset
(Alarm Reset) within 60 seconds, High Priority alarms will sound at the maximum level of 85 dB(A).
There are currently no Low Priority (LP) Alarms.
Summary of Contents for Puritan Bennett 560
Page 1: ...User s Manual Puritan Bennett 560 Ventilator TM...
Page 6: ...This page intentionally blank iv User s Manual...
Page 10: ...viii User s Manual This page is intentionally blank...
Page 14: ...4 User s Manual This page is intentionally blank...
Page 32: ...Safety Information 1 18 User s Manual Figure 1 4 Location of Labels Bottom View 7 13 1 4 3...
Page 98: ...5 10 User s Manual This page is intentionally blank...
Page 128: ...C 2 User s Manual This page is intentionally blank...
Page 134: ...E 4 User s Manual This page is intentionally blank...
Page 148: ...Index Index 6 User s Manual This page intentionally blank...
Page 149: ......