EN-7
Kendall SCD 700 Series
Section II - Battery Operation
The Kendall SCD 700 Series Compression System is designed to operate normally on AC line power or DC battery
power without interruption. There are three Battery Status Indicator LED’s used to represent the charge level of the
battery. Once the Controller is powered on, it may take the system a few seconds to establish communication with
the battery and display the charge level. The battery Indicator shown below is located in the upper right hand corner
of the user interface. See FIGURE 3.
Warning: If the ground integrity of the mains power cable is in question, the device should be operated
on battery power until the ground integrity can be insured.
FIGURE 3
1
3
2
Battery Status Indicators
Unit plugged in and Powered On (Charging)
Battery State
Battery Status 1
Battery Status 2
Battery Status 3
100% charge
Green
Green
Green
67-99% charge
Green
Green
Green (Pulsing)
34-66% charge
Green
Green (Pulsing)
Off
0-33% charge
Green (Pulsing)
Off
Off
Unit not plugged in and Powered On (Operating on Battery)
Battery State
Battery Status 1
Battery Status 2
Battery Status 3
67-100% charge
Green
Green
Green
34-66% charge
Green
Green
Off
< 34% charge
Green
Off
Off
15-40 minutes left*
Amber (Flashing)
Off
Off
< 15 minutes left*
Red (Flashing)
Off
Off