EN-3
Kendall SCD 700 Series
Front Panel Display
1
2
3
6
7
4
5
Item Explanation
Item Explanation
1
Power On Indicator
5
Battery Status Indicators 1-3
2
Power On/Standby Button
6
Right Arrow Button
3
A - Button
7
B - Button
4
AC Power/Battery
Charging Indicator
Section I - General Operating Instructions
Set up
Bed
Foot
Board
• Place the Controller on the footboard. This is done by grasping the device
handle and the top portion of the pivoting bed hook and squeezing to open
the gap. Place it on the foot board so it straddles the foot board and release
the bed clamp. See the figure at right. Ensure its security. If required, alter-
natively, the device can be placed on a horizontal surface appropriate for the
environment, such as on a table, within reasonable proximity to the point of
use. Be sure to allow adequate air flow to the vents located at the power cord
cover and below the tube set connection points.
• The controller can operate with one or two garments attached to the patient.
• Plug the tubing set(s) into the back of the controller. Route the tubing toward
the patient’s limbs, being careful to leave access ways clear and eliminate
tripping hazards.
• Plug the tubes into garment(s) wrapped onto the patient’s limbs.
• Match the left and right ports, marked B and A respectively, with the left and right limbs of the patient. Although
the operation of the controller is not affected, troubleshooting can be easier. Check tubing set(s) for kinking and
secure attachment at the controller and the garment(s).
• Plug the controller power cord into a properly grounded hospital grade receptacle. The blue AC Power Indicator
will illuminate. If no AC Power is accessible, the controller can be run using its own internal battery power.