EN-18
Kendall SCD 700 Series
Test Mode T6 - Performance Test
When in this mode, the user can verify the pump and valve performance, pressure delivery, and the airflow through
the pneumatic circuit. In manufacturing, this test is conducted with known volumes connected to the sleeve s. Then
the inflation cycles run during the test at the low and high pump speeds create backpressures in the volumes that are
measured and used to verify system performance.
• Attach a tubing set connected to leg sleeves wrapped around appropriately sized leg forms.
• Enter Test Access Mode and select Test Access Mode T6.
• Press the B Button to begin the Performance Test.
• The T6 will blink on the display screen until the test is completed.
• After initiating the performance test, the A Leg icon will flash in sync with an audible alarm.
• After the B Button is pressed the A Leg icon will stop flashing, the alarm will stop, and the controller will then go
through a normal inflation cycle on port A with the pump operating at a low speed throughout the cycle.
• Next, the B Leg icon will flash green in sync with an audible alarm.
• After the B Button is pressed, the B Leg icon will stop flashing, the alarm will stop, and the controller will then go
through a normal inflation cycle at B with the pump operating at a high speed throughout the cycle.
• Upon completion the unit beeps and reverts back to Test Access Mode.
Test Mode T7 - Manufacturing Test
Manufacturing Test mode is used in manufacturing with specialized test equipment to ensure proper assembly and
performance. This mode is not intended for use outside of the manufacturing environment.
Test Mode – Alarm History
Alarm History test mode allows the user to access the recent alarm history of a device. It stores the ten most recent
alarms in reverse chronological order. This feature makes diagnosing device problems easy. To view the Alarm History,
enter test access mode and select the Alarm History icon after T7. The Alarm History is shown starting with the most
recent alarm numbered 1. The alarm icon associated with the alarm will be displayed. Each time the Right Arrow
Button is pressed the display will show the next alarm in the reverse chronological order up to 10 alarms Pressing the
button again after the 10th alarm will return the user to the first alarm. If the A or B Button is pressed the controller
will return to Test Access Mode.