Salter AIRE Elite
™
Compressor
Instruction Manual
Part No: 8350-1 / 8352-1 / 8353-1
Thank you for selecting the
Salter AIRE
Elite
Compressor.
Salter Labs is an innovative, industry-leading manufacturer
of respiratory care devices. Please contact your local Salter
Labs dealer for information about additional products.
SAVE THESE INSTRUCTIONS
.
READ ALL INSTRUCTIONS BEFORE USE.
Contents
Important Safeguards………………………………….………………1
Introduction…………………………………………………………………2
Product Description…………………….……………………………..3
Operation……………………………………………………………..…….4
Cleaning……………………………………………..………………………..5
Storage……………………………………………………………………….6
Maintenance………………………………………………………………7
Expected Service Life…………………………………………………8
Troubleshooting…………………………………………………………9
Specifications.……………………………….……………………………10
Accessories….……………………………………………………………11
Appendix A………………………………………………………………12
1. Important Safeguards
NOTE, CAUTION, WARNING, AND SYMBOLS:
Important information is highlighted by using the following:
NOTE
Indicates information that user should pay special attention
to.
CAUTION
Indicates correct operating or maintenance procedures in
order to prevent damage to or destruction of the equipment or other
property.
WARNING
Indicates potential danger that requires correct
procedures or practicesin order to prevent personal injury
.
Symbols:
O
Off, disconnection from the mains
I
On, connection to the mains
~
Alternating Current (AC)
IP2X
Protected against solid foreign objects having a
diameter of 12.5 mm and greater. No protection against
vertically falling water drops, Keep dry !
Attention
Class II
Refer to instruction manual/ booklet
“BF” symbol, indicate this product is according to the
degree of protecting against electric shock for the type
BF equipment.
Temperature limitation
Disposal of Electrical & Electronic Equipment (WEEE):
This product should be handed over to an applicable
collection point for the recycling of electrical and
electronic equipment. For more detailed information
about the recycling of this product, please contact your
local city office, household waste disposal service or the
retail store where you purchased this product.
CAUTION
U.S. Federal Law restricts this device to sale by or on the
order of a physician.
WARNING
To reduce the risk of burns, electrocution, fire or
injury to persons:
1.
Always unplug this product immediately after using.
2.
Do not use while bathing, showering, dish washing, or close to
water sources of any kind.
3.
Do not place or store product where it can fall or be pulled
into a tub or sink.
4.
Do not place in or drop into water or other liquid.
5.
Do not reach for a product that has fallen into water. Unplug
immediately.
6.
This product should never be left unattended when plugged in.
7.
Close supervision is necessary when this product is used by,
on or near children or invalids. Choking accident may result
from a child swallowing a small part that has become detached
from the device or its accessories.
8.
Use this product only for its intended use as described in this
manual. Use this product only under doctor’s direction. Do not
use attachments not recommended by the manufacturer.
9.
Never operate this product if a) it has a damaged cord or plug,
b) it is not working properly, c) it has been dropped or
damaged, d) it has been dropped into water. Return the
product to a specified service center for examination and
repair.
10.
Keep the cord away from heated surfaces.
11.
Never block the air openings of this product or allow objects to
fall or be inserted into the air vent openings or place it on a
soft surface such as bed or couch, where the air openings may
be blocked.
12.
Never use while sleeping or feeling drowsy.
13.
Never drop or insert any object into any opening or hose.
14.
No modification of this equipment is allowed.
15.
Do not modify this equipment without authorization of the
manufacturer.
16.
If this equipment is modified, appropriate inspection and
testing must be conducted to ensure continued safe use of the
equipment.
17.
Do not use in outdoors or operate where aerosol (spray)
products are being used or where oxygen is being
administered in a closed environment such as an oxygen
reservoir.
18.
Do not wrap the power cord around the compressor (main
unit).
19.
Disconnect the power plug by pulling the plug, not by pulling
on the compressor (main unit), or the cord.
20.
If the power cord or plug becomes frayed or otherwise
damaged, do not use.
21.
Do not place heavy objects on the power cord, or bend and
pull the cord harder than necessary. These actions could cause
an electric shock or fire.
22.
Potential allergic reactions to accessible materials used in the
Compressor Nebulizer equipment. If any signs of allergic
reaction or hypersensitivity happen, stop the treatment
immediately, and notify the doctor or nurse.
23.
Potential contact injuries for patients used in the Compressor
Nebulizer equipment. If any contact injuries happen, stop the
treatment, and notify the doctor or nurse.
WARNING
EMC Statement
This equipment has been tested and found to comply with the
limits for medical devices to the EN 60601-1:2007. These limits are
designed to provide reasonable protection against harmful
interference in a typical medical installation. This equipment
generates uses and can radiate radio frequency energy and, if not
installed and used in accordance with the instructions, may cause
harmful interference to other devices in the vicinity. However,
there is no guarantee that interference will not occur in a particular
installation. If this equipment does cause harmful interference to
other devices, which can be determined by turning the equipment
off and on, the user is encouraged to try to correct the interference
by one or more of the following measures:
Reorient or relocate the receiving device.
Increase the separation between the equipment.
Connect the equipment into an outlet on a circuit different from
that to which the other device(s) are connected.
Consult the manufacturer or field service technician for help.
CAUTION
If there is a possibility of electro-magnetic interference
with mobile phones, please increase the distance (3.3m) between
devices or turn off the mobile phone.
2. Introduction
2.1 Intended Use
The
Salter AIRE
Elite
Compressor System is intended to provide a
source of compressed air for aerosol therapy. It is used in
conjunction with a jet (pneumatic) nebulizer to produce medicated
aerosols for inhalation by pediatric and adult patients with
respiratory symptoms.
CAUTION
Aerosolize liquid medication except Pentamidine for
inhalation by the patients. Indications for therapy include asthma,
chronic bronchitis, infection of the upper respiratory tract, chronic
obstructive pulmonary disease (COPD) and other respiratory
disorders in accordance with a medical doctor’s prescription. Except
the usage mentioned above, please do not use this product for any
other purpose. This device can be used with adults or pediatric
patients under physician’s prescription.
2.2 Safety Precaution Instruction
When using this electrical product, especially when children are
present, one should always follow basic safety precautions. Do not
install, maintain or operate this equipment without reading,
understanding and following the proper
Salter AIRE
Elite
Compressor System instruction manual, otherwise injury or
damage may result.
For 120V only-
This appliance has a polarized plug (one blade is
wider than the other). To reduce the risk of electric shock, this plug
is intended to fit into a polarized outlet only one way. If the plug
does not fit fully into the outlet, reverse the plug. If it still does
not fit, contact a qualified electrician. Do not modify the plug in any
way.
3. Product Description
1. Power Switch
5. Power Cord
2. Filter Cap (disposable filter inside) 6. Nebulizer Cup Holder
3. Air-Outlet Connector
7. Cooling Air Openings
4. Integrated Carrying Handle
4. Operation
NOTE
Before initial operation, the nebulizer cup assembly should
be cleaned following instructions described in the “Cleaning”
section.
WARNING
Before connecting the power cord, make sure the I/O
(ON/OFF) switch is in the O (OFF) position.
The plug is also served to disconnect the device. Do not
position the equipment so that it is difficult to operate
the disconnecting device.
4.1 Daily Use Operation
CAUTION
The Salter AIRE Elite Compressor System is designed for
intermittent use only. Do not operate it continuously for more than
30 minutes for a single use without turning it off and following a
cooling period for least 30 minutes.
1.
Before each use inspect the
Salter AIRE
Elite
Compressor
and nebulizer cup assembly for damage or wear, replace as
needed.
2.
Place the
Salter AIRE
Elite
Compressor on a table or other
flat stable surface. Be sure you can easily reach the controls
when seated. Do not use this device on the floor.
3.
With the power switch in the O (OFF) position, plug the power
cord into an appropriate electrical wall outlet.
4.
Connect one end of the tubing to the compressor air-outlet
connector.
5.
Assemble the nebulizer cup and add indicated medication to
the nebulizer’s cup before use.
6.
Attach the other end of the tubing into the air-inlet connector
founded at the bottom of the nebulizer cup.
7.
Turn on the
Salter AIRE
Elite
Compressor by pressing the
power switch to the I (ON) position and begin treatment.
8.
If treatment needs to be interrupted, simply press power
switch to O (OFF) position.
9.
When the treatment is complete, turn off the compressor by
pressing the power switch to O (OFF) position and unplug the
unit from the electrical outlet.
Equipment is not suitable for use in the presence of a
flammable anesthetic mixture with air or with oxygen or
nitrous oxide. This device has no AP or APG protection.
CAUTION
During the treatment, the patient should not touch the
outer case due to expected
rise in unit temperature.
4.2 Safety Overload
NOTE
Do not exceed 30 minutes of continuous operation.
1.
The motor of this device has a built-in thermal overload
protector. Should the motor overheat, the protector will
automatically shut off the motor. Should this occur, turn the
I/O (ON/OFF) switch to the O (OFF) position and allow the
motor to cool down for approximately 30 minutes before turn
it on again.
2.
If the overload protector shuts off the motor frequently, you
may have an unstable voltage situation.
3.
If the unit shuts down and cannot restart, it may need to be
replaced. Call your provider immediately.
5. Cleaning
5.1 Compressor Outer Case Cleaning
WARNING
Electric shock hazard. Do not remove outer case of
this unit. All disassembly and maintenance of this unit must be
done by a qualified service technician. Refer servicing to
qualified service personnel.
WARNING
This unit does not require oil. Do not attempt to
lubricate any internal parts.
WARNING
Unplug unit before cleaning. Do not submerge in
water for cleaning.
1.
Wipe the main unit with a damp cloth every few days to keep
it dust-free.
2.
Do not use any powdered type cleaners or soap. Do not
submerge the unit into water.
5.2. Nebulizer Cup Cleaning
Clean the nebulizer after each use. Refer to the cleaning
instructions supplied with your nebulizer
800987 Rev D (Side 1)
Printed on 15 April 2015