Planar
TR1300/1,
TR5048,
TR7530
Programming
Manual
SCPI
228
SYST:ERR?
SYSTem:ERRor[:NEXT]?
Description
Reads
out
the
error
message
of
the
error
occurred
when
executing
the
SCPI
commands,
from
the
FIFO
(First
In
First
Out)
error
queue
stored
in
the
instrument.
The
read
out
error
is
deleted
from
the
error
queue.
The
*CLS
command
clears
the
error
queue.
The
maximum
size
of
the
queue
is
100
messages
(query
only).
Query
Response
<numeric>,
<string>
Where:
<numeric>
error
code
<string>
error
message
If
there
is
no
error
in
the
queue,
"0,
No
error"
is
read
out.
Equivalent
Softkeys
None
SYST:LOC
SYSTem:LOCal
Description
Sets
the
instrument
to
the
local
operation
mode,
when
all
the
keys
on
the
front
panel,
mouse
and
the
touch
screen
are
active
(no
query).
Related
Commands
SYST:REM
SYST:RWL
Equivalent
Softkeys
None